Gas System


Gas Rack
HIGH SPEED GAS CHROMATOGRAPHY
IN-LINE TOTAL GAS (0 - 100% METHANE)
VORTEX GAS TRAP SYSTEM ALLOWING CONSTANT FLOW / VOLUME.
Explosion Proof Electrical or compressed air gas trap in-house designed STAINLESS STEEL to prevent corrosion).
 
GAS TRAP
Our Gas Trap allows the continuous extraction of gases from drilling mud so that they can be analyzed.
Internal baffles and a 3 blade agitator were designed to reduce the effect of immersion level variations on gas extraction yields.
Compressed Air or ATEX Electrical Motor

Characteristics 
Diameter    154 mm
Mud entry diameter  45 mm
Height of mud exit port  152 mm
Body height  310 mm
Mud Flow  15 to 25 l/min
Gas suction rate   2 to 4 l/min
Rotation speed    1800 rpm
Construction materials  316L stainless steel
PVC gas line is 4 mm ID  
 

FID CHROMATOGRAPH 8900

The Model 8900 Well Logging Gas Analyzer combines the selectivity of gas chromatography with the sensitivity, broad dynamic range and hydrocarbon selectivity of a flame ionization detector (FID) providing fast analysis (56 seconds) of C1-C5 hydrocarbons.

The 8900 WL Analyzer uses a dual column configuration with timed backflush to vent which is used to strip of moisture and C5 hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection, a fixed sample volume is carried to the pre-cut column. Backflush of the pre-cut column is timed from sample injection so that only C1-C5 hydrocarbons are eluted from the pre-cut column to the analytical column.

Gas Chromatograph (Skychrom) Control View  
 
Analysis Time : 56 seconds
Detector : FID
Column Proprietary : Micropacked Column
Oven Temperature : 80°C
Carrier Hydrogen : 32cc/min typical
FueL Hydrogen : 8cc/min typical
Combustion Air : 200cc/min typical
 

Analysis of C1-nC5 Hydrocarbons including Propylene for Gas Referencing
Automatic Restart for Unattended Operation
Automatic Zeroing for Long-Term Stability
Automatic Ignite of FID
Automatic Hydrogen & Combustion Air Shutoff for Maximum Safety
Automatic or Manual Calibration to a Known Standard
Electronic Support Gas Regulation

 
 

FID TOTAL GAS DETECTOR 9000
LOWER detection : 10 ppm for methane
LINEAR RANGE : from 100 ppm to 100% for methane
ACCURACY : ± 5% (relative) within dynamic measurement ranges of 100 ppm to 20%, 2.54% relative from 20% to 50%; 2% relative from 50 to 100%
PRECISION : 2% of Measured Value
ANALYSIS TIME : 5 Seconds
CARRIER Hydrogen : 35cc/min
COMBUSTION AIR : 175cc/min typical
OVEN TEMPERATURE : 80°C

 


OTHER GAS DETECTION


H2S monitorING.

3 sensors to be installed on rig, one sensor on gas line.
External (siren) and Internal audible alarms.
External visible alarm (gyroscope) - YELLOW

COMBUSTIBLE GAS MONITORING
3 CH4 sensors installed on rig structure.
External (siren) and Internal audible alarms
External visible alarm (gyroscope) - BLUE

CO2
Optional CO2 detector recorded on line.