Gas System
Gas Rack |
HIGH SPEED GAS CHROMATOGRAPHY IN-LINE TOTAL GAS (0 - 100% METHANE) VORTEX GAS TRAP SYSTEM ALLOWING CONSTANT FLOW / VOLUME. Explosion Proof Electrical or compressed air gas trap in-house designed STAINLESS STEEL to prevent corrosion). |
Our Gas Trap allows the continuous extraction of gases from drilling mud so that they can be analyzed.
Internal baffles and a 3 blade agitator were designed to reduce the effect of immersion level variations on gas extraction yields.
Compressed Air or ATEX Electrical Motor
Characteristics
Diameter | 154 mm | ||
Mud entry diameter | 45 mm | ||
Height of mud exit port | 152 mm | ||
Body height | 310 mm | ||
Mud Flow | 15 to 25 l/min | ||
Gas suction rate | 2 to 4 l/min | ||
Rotation speed | 1800 rpm | ||
Construction materials | 316L stainless steel | ||
PVC gas line is 4 mm ID | |||
FID CHROMATOGRAPH 8900
The Model 8900 Well Logging Gas Analyzer combines the selectivity of gas chromatography with the sensitivity, broad dynamic range and hydrocarbon selectivity of a flame ionization detector (FID) providing fast analysis (56 seconds) of C1-C5 hydrocarbons.
The 8900 WL Analyzer uses a dual column configuration with timed backflush to vent which is used to strip of moisture and C5 hydrocarbons. A pre-cut column is used in series with the analytical column. At sample injection, a fixed sample volume is carried to the pre-cut column. Backflush of the pre-cut column is timed from sample injection so that only C1-C5 hydrocarbons are eluted from the pre-cut column to the analytical column. |
|
Gas Chromatograph (Skychrom) Control View |
Detector : FID
Column Proprietary : Micropacked Column
Oven Temperature : 80°C
Carrier Hydrogen : 32cc/min typical
FueL Hydrogen : 8cc/min typical
Combustion Air : 200cc/min typical
Analysis of C1-nC5 Hydrocarbons including Propylene for Gas Referencing
Automatic Restart for Unattended Operation
Automatic Zeroing for Long-Term Stability
Automatic Ignite of FID
Automatic Hydrogen & Combustion Air Shutoff for Maximum Safety
Automatic or Manual Calibration to a Known Standard
Electronic Support Gas Regulation
FID TOTAL GAS DETECTOR 9000
LOWER detection : 10 ppm for methane
LINEAR RANGE : from 100 ppm to 100% for methane
ACCURACY : ± 5% (relative) within dynamic measurement ranges of 100 ppm to 20%, 2.54% relative from 20% to 50%; 2% relative from 50 to 100%
PRECISION : 2% of Measured Value
ANALYSIS TIME : 5 Seconds
CARRIER Hydrogen : 35cc/min
COMBUSTION AIR : 175cc/min typical
OVEN TEMPERATURE : 80°C
OTHER GAS DETECTION
H2S monitorING.
3 sensors to be installed on rig, one sensor on gas line.
External (siren) and Internal audible alarms.
External visible alarm (gyroscope) - YELLOW
3 CH4 sensors installed on rig structure.
External (siren) and Internal audible alarms
External visible alarm (gyroscope) - BLUE
CO2
Optional CO2 detector recorded on line.